Category
Related Products
|
Order Code |
Product Name |
|
H-06-02-M01 |
23L Pre-vacuum Pressure Steam Autoclave |
Application
Centralized disinfection of hospital stomatology, ophthalmology, beauty hospital, operating room, supply room
Feature & Technical specification
1. CE conformity. Heating method: electric heating
2. Power: 1.7kVA
3. Power supply: AC220V, 50Hz
4. Dimensions (L×W×H): 715×513×425mm
5. Inner tank size (diameter x depth): 246 x 450mm
6. Effective volume: 23L
7. Weight: 49kg
8. Energy consumption: about 1.5 kWh per pot
9. Water tank volume: 2 L
10. Water consumption: 0.5-1L (depending on the load and type)
11. Discharge volume: 0.5-1L (except for a very small part remaining on the sterilized items, the rest are all excluded)
12. Sterilization chamber material: The sterilization chamber is made of 304 anti-corrosion stainless steel
13. Outer box material: ABS material
14. Working environment: 5℃~40℃
15. Working pressure: -80kPa~210kPa
16. Humidity ≥80% during sterilization
17. The highest pressure resistance value: -99kPa~280kPa
18. Temperature during sterilization: 121℃/134℃
19. Temperature display: LCD display
20. Microcomputer control technology: The sterilizer uses microprocessor intelligent control, automatic door locking, man-machine interface, easy operation, safe and reliable
21. Vacuum method: pulsed vacuum multiple times to ensure faster heating and more thorough sterilization
22. Safety valve: when the pressure of the sterilizer reaches 250kPa, the safety valve will automatically open to relieve the pressure to ensure safety
23. Overpressure protection program: when the pressure of the sterilization chamber reaches 220kPa, the equipment can automatically exhaust steam to balance the pressure. When the pressure of the sterilization chamber reaches 230kPa, the program will terminate the operation and automatically release the pressure
24. Air leakage: When the sterilizer is tested at each stage, the pressure rise rate should not exceed 0.13kPa/min
25. Water quality monitoring: with automatic water quality monitoring function
26. Sterilization chamber dynamic pressure: The pressure change in any 2s interval during the sterilizer cycle should not exceed 1000kPa/min
27. Vacuum pump: use leather bowl double-head oil-free vacuum pump
28. Pressure sensor: The built-in pressure sensor automatically detects the pressure in the cavity
29. Temperature sensor: The built-in temperature sensor automatically detects the temperature in the cavity
30. Printer: Micro-printer is used to print out every minute data in the whole process of sterilization, which can be stored permanently
31. Sterilization mode: bare instruments, encapsulated instruments, fabric hollow, rubber products, BD test, vacuum test
32. Automatic drying: powerful vacuum drying, suitable for drying requirements of different items; residual humidity of equipment <0.2%, residual humidity of dressing <1%
33. Internal steam discharge: built-in condensate collector, waste steam is discharged after condensation by the condenser, and no waste steam is discharged to the outside, which is safer and cleaner
34. Sealing structure: The sealing ring is made of silicone, and the designed service life is 2 years (1200 times) without aging and deformation
35. Wearing parts: sealing ring
36. Alarm system: with high and low temperature alarm system
Contact Information
E-mail:
Tel:
Address : Room 2512, Unit 1, Huaqiao Square, No.1, Daqiao North Road, Jiangbei New District, Nanjing, China